Experimental Station 8-1

BL8-1 is equipped with a post-monochromator beam-splitting mirror to allow for two sample chamber locations. BL8-1a accommodates roll-on user chambers. BL8-1b is under modification to provide energy range from 2 eV to 10 eV. The mirror cooling system was upgraded in FY04 and the BL8-1 stations are SPEAR3 500 mA compatible. Stations 8-1a and 8-1b cannot be used simultaneously. This experimental station is NOT open to users.

Status
Closed
Supported Techniques
Photoemission spectroscopy
Soft x-ray absorption spectroscopy
Main Scientific Disciplines
Catalysis
Materials Sciences
Physics
Beam Line Specifications

Source

Bending Magnet, 8.0 mrad acceptance

  Energy Range Resolution ΔE/E Spot Size Flux (ph/sec)
Focused 15 - 185 eV ~ 1 X 10-3 ≥ 0.1 X 0.1 mm2 1 X 1012

 

Optics

  • M0: Cylinder SiC
  • M1: Spherical
  • Monochromator: 6 m TGM

Absorption

N/A

Instrumentation
  • 15 - 185 eV TGM.
  • Available for “Roll-up” Systems.
  • Facility Hemispherical Analyzer,
  • CMA-Based System
  • User Chambers on VUV Lines
Sample Environment

N/A

Data Acquisition and Analysis

N/A

Beam Line Phone
650-926-5281
Beam Line Contact
Beam Line Engineering Notes